Safeway, in Cooperation with Trividia Health Inc., is Voluntarily Recalling Safeway TRUEtrack and Signature Care TRUEtest Blood Glucose Test Strips Due to a Packaging Defect That May Affect Testing Accuracy
PLEASANTON, Calif., July 8, 2016 /PRNewswire/ -- Safeway, in cooperation with Trividia Health, Inc., formerly Nipro Diagnostics, is voluntarily recalling all lots of Safeway TRUEtrack Blood Glucose Test Strips and Signature Care TRUEtest Blood Glucose Test Strips.
The products, manufactured by Trividia Health, were sold at all Safeway, Vons, Pavilions, Randalls, Tom Thumb, Carrs and Pak 'N Save stores nationwide. Trividia Health manufactures blood glucose test strips for more than 100 retailers.
Trividia Health reports that some blood glucose test strip lots, packaged during a specific timeframe, may include open, unsealed test strip vials within the sealed test strip vial boxes. Exposure to an open environment for long periods can affect test strip performance. If a user does not notice an open test strip vial upon opening the sealed test strip vial box, and then uses the test strips to measure blood glucose, the meter may provide incorrect low blood glucose results.
The recall of Safeway TRUEtrack and Signature Care TRUEtest Test Strips includes the product sizes and date codes listed below.
Product Name |
Size |
UPC Code (On bottom of box) |
Lot Code or Expiry Date (On the back panel) |
Signature Care TRUEtest Blood |
25 count |
21130-77126 |
PS2219 - Use By 1/31/2018 |
Signature Care TRUEtest Blood |
50 count |
21130-77124 |
PS2278 – USE BY 4/14/2018 PS2303 – USE BY 4/30/2018 PS2330 – USE BY 5/22/2018 PS2365 – USE BY 5/31/2018 PS2381 – USE BY 6/30/2018 |
Signature Care TRUEtest Blood |
100 count |
21130-77125 |
PS2280 – USE BY 4/14/2018 PS2314 – USE BY 5/14/2018 PS2371 – USE BY 6/16/2018 PS2379 – USE BY 6/30/2018 PS2408 – USE BY 7/16/2018 |
Safeway TRUEtrack Blood |
50 count |
21130-77111 |
RS4655 – USE BY 10/17/2017 RS4673 – USE BY 10/31/2017 RS4708 – USE BY 12/16/2017 |
Safeway TRUEtrack Blood |
100 count |
21130-77112 |
RS4567 – USE BY 5/31/2017 RS4704 – USE BY 11/30/2017 RS4716 – USE BY 12/25/2017 |
These products are sold over-the-counter near the pharmacy or in the medication aisle.
No injuries or illnesses have been reported in relation to the Safeway recall. This recall is being issued out of an abundance of caution.
Customers are asked to discard the product or return it to their local store for a full refund. Customers who have questions can contact Trividia Health at www.trividiahealth.com/product-notice or call Stericycle toll-free at 1-888-835-2723. Customers can also contact Safeway at 1-877-276-9637.